Abbe refractometer AR2008

4.867,00 

Digital display refractive index and %Brix: resolution (0.0001 nD); measurement range (nD1.3000-1.7200)

Description

The AR2008 is an Abbe refractometer in which the sample is placed between a measuring and illumination prism. The device can measure the refractive index or Brix value and display the measured value together with the temperature on the LED display.

With the serial RS-232 interface, the AR2008 can be integrated into existing systems. The measured value and temperature with date and time can be transmitted directly to a printer, network or PC – a decisive advantage for quality management and traceability.

To ensure temperature control, the refractometer can be connected to a circulation thermostat. These devices can be used in a wide range of applications: e.g. in the chemical, petrochemical, oil and grease industries, the food industry, in the pharmaceutical environment and in research & education.

Basic functions

  • Measurement of liquid and viscous samples, regardless of their viscosity
  • Measurement of films possible and of solid samples that have at least one smooth side
  • Wide measurement range from nD 1.3000-1.7200 and 0-95 %Brix
  • Digital display of measured data
  • Integrated thermometer
  • Hose connections for temperature control with a circulation thermostat
  • Automatic temperature compensation for the BRIX scale can be switched on as an option
  • Robust housing and easy handling
  • Serial interfaces for coupling with PC or printer (RS-232, RS-422)

Application Abbe refractometer

The AR2008 has a larger and more robust design, making it ideal for demanding laboratory and production environments. The high-quality optics provide a clearly defined bright/dark line so that measured values can be recorded more quickly and reliably. The digital scale reading also ensures precise, error-free results, even with changing light conditions or frequent user changes.

Modified atmosphere packaging ensures the quality of baby food

Application example food industry

  • Valid norms/standards: AOAC, OIML, ASTM, GLP, ICUMSA
  • Application: Quality and purity control of raw materials and end products (convenience products, dairy products, baby food, egg products, oils). Ideal for high sample volumes.
  • Quick measurement and determination of sugar concentration (jams, honey, syrup, seasoning sauces, mustard, mayonnaise)
  • Advantage: easy handling, easy cleaning. Good image quality for the bright/dark line, digital printout option.

KRUESS-Optronic-Abbe refractometer-Hospitals-Pharmacies

Application example hospitals and pharmacies

  • Valid norms/standards: Pharmacopoeias (BP,JP, Ph. Eur.), GLP
  • Application: Outgoing goods inspection or analysis of raw materials. Control of drugs according to Pharmacopoeia (infusion solutions, blood serums, pills, pastes, dialysis preparations) or analysis of body secretions
  • Advantage: High reliability Scale can be read digitally. Availability of the BRIX scale, possibility of temperature compensation. Provides an RS-232 interface.


Overview samples


Download application areas

Technical data

TABLE

Spezifikation Abbe-Refraktometer AR2008 Modell
Messbereich nD 1,3000–1,7200
0–95 %Brix
Temperierbereich 1) Nachrüstbar
Messgenauigkeit 2) nD ±0,0002
±0,1 %Brix
Temperiergenauigkeit 1) Nachrüstbar
Auflösung nD 0,0001
0,1 %Brix
Wellenlänge 589 nm
Temperaturmessgenauigkeit ±0,3 °C
Temperaturmessauflösung 0,1 °C
Temperierung 1) Nachrüstbar
Skalen Brechungsindex (nD), Konzentration von Saccharose (%Brix)
Justierung Über Stellschraube
Probenberührende Teile Optisches Flintglas (SF4)
Zusätzliche Funktionen Optionale Temperaturkompensation für %Brix nach ICUMSA
Bedienung Triebknopf
Auswahltasten
Schnittstellen 1 x RS–232, 1 x RS–422
Maße (B x H x T) 120 mm x 290 mm x 250 mm
Gewicht 5 kg
Betriebsspannung 115/230 V, umschaltbar
Leistungsaufnahme (Max.) 40 W

¹) Erfordert Umwälzthermostat PT31
²) Bei Normalbedingungen für Messung des Brechungsindex (λ = 589 nm, 20 °C, 1013 hPa, 50 % rel. Feuchte)

Standards & guidelines

  • The AR2008 is also used in industry because it can be used where standards, precise measurement procedures and secure data transmission are required.
  • Companies that need to document measured values seamlessly benefit from the Abbe refractometer’s serial RS-232 interface. It enables the direct transmission of measured values, temperature, date and time to a PC, network or printer.
  • We have compiled an overview of relevant standards for refractometer measurements for your guidance. This helps with the selection of the necessary equipment.


Overview of standards and guidelines

Documents

Titel Herunterladen
Produktinformation Refraktometer DR7000-Serie HerunterladenVorschau
Broschüre Refraktometer HerunterladenVorschau
Handrefraktometer Skalen und Messbereiche HerunterladenVorschau
Refraktometer Anwendungsbereiche HerunterladenVorschau
Refraktometer Proben Übersicht in englisch HerunterladenVorschau
Refraktometer Normen Übersicht in englisch HerunterladenVorschau
Reinigungstipps-Refraktometer HerunterladenVorschau
Prozessrefraktometer Anforderungsprofil HerunterladenVorschau


Calibration standards

Consumables

Accessories


Click here

Weight 6,2 kg

Withdraw from contract

Scroll to Top